10
OUTPut[1|2]:IMPedance
This command sets the output load impedance for the specified channel. The specified
value is used for amplitude, offset, and high/low level settings. You can set the impedance
to any value from 1 Ω to 10 kΩ with a resolution of 1 Ω. The default value is 50 Ω.
The query returns the current load impedance setting in ohms. If the load impedance is
set to INFinity, the query returns “9.9E+37”.
Syntax
OUTPut[1|2]:IMPedance {<ohms>|INFinity|MINimum|MAXimum}
OUTPut[1|2]:IMPedance? {MINimum|MAXimum}
Arguments
<ohms>::=<NR3>[<units>]
where:
<units>::=OHM
INFinity sets the load impedance to >10 kΩ.
MINimum sets the load impedance to 1 Ω.
MAXimum sets the load impedance to 10 kΩ.
Returns
<ohms>::=<NR3>
Examples
OUTPut1:IMPedance MAXimum
sets the CH1 load impedance to 10 kΩ.
OUTPut[1|2][:STATe]
This command sets or query the instrument output state for the specified channel.
Syntax
OUTPut[1|2][:STATe] {ON|OFF|<NR1>}
OUTPut[1|2][:STATe]?
Arguments
ON or <NR1>≠0 enables the instrument output.
OFF or <NR1>=0 disables the instrument output.
Returns
<NR1>