<seconds>::=<NRf>[<units>]
where:
<NRf> is the sweep time in seconds.
<units>::=[ns|us|ms|s]
Returns
<seconds>
Examples
SOURce1:SWEep:TIME 100ms
sets the CH1 sweep time to 100 ms.
[SOURce[1|2]]:VOLTage[:LEVel][:IMMediate]:OFFSet
This command sets or queries the offset level for the specified channel.
Syntax
[SOURce[1|2]]:VOLTage[:LEVel][:IMMediate]:OFFSet {<voltage>|MINimum|MAXimum}
[SOURce[1|2]]:VOLTage[:LEVel][:IMMediate]:OFFSet? {MINimum|MAXimum}
Arguments
<voltage>::=<NR3>[<units>]
where:
<NR3> is the offset voltage level.
<units>::=[mV|V]
Returns
<voltage>
Examples
SOURce1:VOLTage:LEVel:IMMediate:OFFSet 500mV
sets the CH1 offset level to 500 mV.
[SOURce[1|2]]:VOLTage[:LEVel][:IMMediate][:AMPLitude]
This command sets or queries the output amplitude for the specified channel. The
amplitude resolution is 1 mVpp or four digits.
Syntax
[SOURce[1|2]]:VOLTage[:LEVel][:IMMediate][:AMPLitude]