EasyManua.ls Logo

VOLTCRAFT FG-1602 - System:beeper[:Immediate] (no Query Form); System:beeper:state

VOLTCRAFT FG-1602
57 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
47
{<amplitude>|MINimum|MAXimum}
[SOURce[1|2]]:VOLTage[:LEVel][:IMMediate][:AMPLitude]? {MINimum|MAXimum}
Arguments
<amplitude>::=<NR3>[<units>]
where:
<NR3> is the output amplitude.
<units>::=[mVpp|Vpp]
Returns
<amplitude>
Examples
SOURce1:VOLTage:LEVel:IMMediate:AMPLitude 1Vpp
sets the CH1 output amplitude to 1 Vpp.
SYSTem:BEEPer[:IMMediate] (No Query Form)
This command causes the instrument to beep immediately. This command is only
available when the beeper is enabled.
Syntax
SYSTem:BEEPer[:IMMediate]
Arguments
None
Examples
SYSTEM:BEEPER
causes a beep.
SYSTem:BEEPer:STATe
The SYSTem:BEEPer:STATe command sets the beeper ON or OFF.
The SYSTem:BEEPer:STATe? command returns “0” (OFF) or “1” (ON).
When the beeper is set to ON, the instrument will beep when an error message or a
warning message is displayed on the screen. The instrument does not beep when an
error or warning caused by remote command execution.