EasyManua.ls Logo

VOLTCRAFT FG-1602 - [Source[1|2]]:Frequency:start; [Source[1|2]]:Frequency:stop

VOLTCRAFT FG-1602
57 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
25
where:
<NRf> is the frequency span.
<units>::=[Hz|kHz|MHz]
Returns
<frequency>
Examples
SOURce1:FREQuency:SPAN 900 kHz
sets the CH1 frequency span to 900 kHz.
[SOURce[1|2]]:FREQuency:STARt
This command sets or queries the start frequency of sweep for the specified channel. This
command is always used with the [SOURce[1|2]]:FREQuency:STOP command. The
setting range of start frequency depends on the waveform selected for sweep.
Syntax
[SOURce[1|2]]:FREQuency:STARt {<frequency>|MINimum|MAXimum}
[SOURce[1|2]]:FREQuency:STARt? {MINimum|MAXimum}
Arguments
<frequency>::=<NRf>[<units>]
where:
<NRf> is the start frequency.
<units>::=[Hz|kHz|MHz]
Returns
<frequency>
Examples
SOURce1:FREQuency:STARt 10kHz
sets the sweep start frequency of CH1 to 10 kHz.
[SOURce[1|2]]:FREQuency:STOP
This command sets or queries the stop frequency of sweep for the specified channel. This
command is always used with the [SOURce[1|2]]:FREQuency:STARt command. The
setting range of stop frequency depends on the waveform selected for sweep.

Table of Contents