40
Syntax
[SOURce[1|2]]:PULSe:DCYCle {<percent>|MINimum|MAXimum}
[SOURce[1|2]]:PULSe:DCYCle? [MINimum|MAXimum]
Arguments
<percent>::=<NR2>[<units>]
where:
<NR2> is the duty cycle.
<units>::=PCT
Returns
<percent>
Examples
SOURce1:PULSe:DCYCle 80.5
sets the duty cycle of the pulse waveform on CH1 to 80.5%.
[SOURce[1|2]]:PULSe:TRANsition[:LEADing]
This command sets or queries the leading edge time of pulse waveform.
Syntax
[SOURce[1|2]]:PULSe:TRANsition[:LEADing] {<seconds>|MINimum|MAXimum}
[SOURce[1|2]]:PULSe:TRANsition[:LEADing]?[MINimum|MAXimum]
Arguments
<seconds>::=<NRf>[<units>]
where
<NRf> is the leading edge time of pulse waveform.
<units>::=[ns|us|ms|s]
Returns
<seconds>
Examples
SOURce1:PULSe:TRANsition:LEADing 200ns
sets the CH 1 leading edge time to 200 ns.