WM 68331 04/2021
LIFE-BASE EN 125
7 Disassembly
4. Remove the portable unit from the device.
Result The device is removed from the portable unit.
7.5.3 Removing the device from LIFE-BASE light
and LIFE-BASE III
Applies to MEDUMAT Standard
a
MEDUMAT Easy
CPR
MODUL
The removal of MEDUMAT Transport from LIFE-BASE light is
described in a separate section (see "7.5.5 Removing the device
from LIFE-BASE 1 NG and LIFE-BASE 4 NG (with protective
hoops)", page 128).
Required auxiliary
equipment
Screwdriver with crosshead PH1
Requirement Disassembly is prepared (see "7.5.1 Preparing disassembly",
page 123).
1. Place the portable unit and device with the control panel
membrane facing downwards.
2. Undo 2 screws 100 and 2 protective disks 104 from
the portable unit.
100
MEDUMAT Easy
CPR
104
LIFE-BASE light