32 03/2016 – UM-MT-EN-01
3.6.1 North America
Requires at least 115 V
from wall outlet.
If voltage falls 10% below
115 V, treadmill will shut off
and reset.
Requires at least 208/230 V
from wall outlet.
If voltage falls 10% below
208/230 V, treadmill will
shut off and reset.
Requires at least 208/230 V
from wall outlet.
If voltage falls 10% below
208/230 V, treadmill will
shut off and reset.
20 Amp
Dedicated line required
(cannot share neutral line)
20 Amp
Dedicated line required
(cannot share neutral line)
30 Amp
Dedicated line required
(cannot share neutral line)
NEMA 5-20 R
(dedicated circuit required)
NEMA 6-20 R
(dedicated circuit required)
NEMA L6-30
(twist lock, 208-230 VAC, 30
Amp dedicated circuit required)
Standard 3-prong, hospital
grade plug (NEMA 5-20 P)
Will only fit a NEMA 5-20 R
outlet
3-prong plug
(NEMA 6-20 P)
Will only fit a NEMA 6-20 R
outlet
3-prong twist plug
(NEMA L6-30 P)
Will only fit a NEMA L6-30 R
outlet
Hospital-Grade
Low Leakage
For grounding reliability,
only connect to proper
receptacle marked “Hospital
Grade” when using for
medical use.