pH value Multi 3620 IDS
37 ba77169e01 04/2016
Continuing with
three- to five-point
calibration
16 Set the nominal buffer value for the measured temperature with
<><>.
17 Accept the calibration value with <ENTER>.
The calibration display for the next buffer appears (voltage display).
18 If necessary, finish the calibration procedure as a two-point calibration
with <M>.
The calibration record is displayed.
15.03.2016 08:00
ConCal
19 Thoroughly rinse the IDS-pH sensor with deionized water.
20 Immerse the IDS-pH sensor in the next buffer solution.
21 When measuring without temperature sensor:
Measure the temperature of the buffer manually and enter it with
<><>.
22 Start the measurement with <ENTER>.
The measured value is checked for stability (stability control).
The [AR] status indicator is displayed. The measured parameter
flashes.
23 Wait for the measurement with stability control to be completed or ter-
minate the stability control and take over the calibration value with
<ENTER>.
The pH value of the buffer solution is displayed.
15.03.2016 08:00
ConCal

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the wtw Multi 3620 IDS and is the answer not in the manual?

wtw Multi 3620 IDS Specifications

General IconGeneral
Brandwtw
ModelMulti 3620 IDS
CategoryMeasuring Instruments
LanguageEnglish

Related product manuals