pH value Multi 3620 IDS
37 ba77169e01 04/2016
Continuing with
three- to five-point
calibration
16 Set the nominal buffer value for the measured temperature with
<><>.
17 Accept the calibration value with <ENTER>.
The calibration display for the next buffer appears (voltage display).
18 If necessary, finish the calibration procedure as a two-point calibration
with <M>.
The calibration record is displayed.
19 Thoroughly rinse the IDS-pH sensor with deionized water.
20 Immerse the IDS-pH sensor in the next buffer solution.
21 When measuring without temperature sensor:
Measure the temperature of the buffer manually and enter it with
<><>.
22 Start the measurement with <ENTER>.
The measured value is checked for stability (stability control).
The [AR] status indicator is displayed. The measured parameter
flashes.
23 Wait for the measurement with stability control to be completed or ter-
minate the stability control and take over the calibration value with
<ENTER>.
The pH value of the buffer solution is displayed.