Settings Multi 3620 IDS
83 ba77169e01 04/2016
Setting
menuTetraCon 925
Menu item Possible
setting
Explanation
Calibration /
Calibration record
- Displays the calibration record of the
last calibration.
Calibration /
Calibration data
storage / Display
- Displays the last calibration records
(max. 10)
Calibration /
Calibration data
storage / Output to
USB flash drive or
printer
- Outputs the stored calibration data
to a connected USB memory device/
USB printer)
Calibration /
Calibration data
storage / Output to
RS232/USB
- Outputs the calibration data storage
to the interface
Calibration /
Calibration interval
1 ... 150 ...
999 d
Calibration interval for the IDS con-
ductivity sensor (in days).
The meter reminds you to calibrate
regularly by the flashing sensor sym-
bol in the measuring screen.
Type Measuring cell used
Cal Measuring cells whose cell constant
is determined by calibration in the
KCL control standard solution.
Calibration range:
0.450 to 0.500 cm
-1
The currently valid cell constant is
displayed in the status line.
man Freely adjustable cell constant in the
range 0.450 ... 0.500 cm
-1
.
Man. cell const. 0.450 ...
0.475 ...
0.500 cm
-1
Display and setting options for the
cell constant.
Temp. comp. (TC)
/ Method
nLF
Lin
off
Procedure for temperature compen-
sation (see section 8.2).
This setting is only available for the
measured parameters, ϰ and ρ.
Temp. comp. (TC)
/
Linear coeff.
0.000 ...
2.000 ...
10.000 %/K
Coefficient of the linear temperature
compensation.
This menu item is only available
when the linear temperature com-
pensation is set.