Settings Multi 3620 IDS
89 ba77169e01 04/2016
The sensor settings are reset under the Reset menu item in the menu for cali-
bration and measurement settings. To open the settings, display the required
measured parameter in the measured value display and press the <ENTER>
key.
Conductivity The following settings for conductivity measurements are reset to the default
settings with the Reset function:
The sensor settings are reset under the Reset menu item in the menu for cali-
bration and measurement settings. To open the settings, display the required
measured parameter in the measured value display and press the <ENTER>
key.
Turbidity The following settings for the turbidity measurement are reset to the default
condition with the Reset function:
Setting Default settings
Kal.-Intervall 180 d
Check interval 60 d
Measured parameter D.O. concentration
Relative slope (S
Rel
)1.00
Salinity (value) 0.0
Salinity (function) Off
Setting Default settings
Kal.-Intervall 150 d
Measured parameter
ϰ
Cell constant (c) Depending on the connected
measuring cell:
0.475 cm
-1
(calibrated)
0.475 cm
-1
(set)
0.100 cm
-1
Temperature compensation nLF
Reference temperature 25 °C
Temperature coefficient (TC) of
the linear temperature compensa-
tion
2.000 %/K
TDS factor 1.00

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the wtw Multi 3620 IDS and is the answer not in the manual?

wtw Multi 3620 IDS Specifications

General IconGeneral
Brandwtw
ModelMulti 3620 IDS
CategoryMeasuring Instruments
LanguageEnglish

Related product manuals