OxiTop
®
Control OC 100 Commissioning
ba31114e06 06/2012 11
3 Commissioning
3.1 Scope of delivery
Controller OxiTop
®
OC 100
Battery holder with batteries (3 x Mignon, AA, AM3, LR6)
Operating manual
Additional scope of delivery: depending on set composition
3.2 Power supply
The controller OxiTop
®
OC 100 and measuring head OxiTop
®
-C are both
battery-powered.
Additionally, the controller has a data backup battery. This battery ensures that
your measuring data and instrument settings are stored in the controller
OxiTop
®
OC 100.
3.2.1 Batteries
3.2.2 Battery status signals
The OxiTop
®
-OC 100 controller and the OxiTop
®
-C measuring heads are pow-
ered by batteries. A low battery status of the controller or measuring head is
automatically indicated by a display message and a signal tone after the instru-
ment is switched on (see section 10.2 D
ISPLAY MESSAGES ON THE CHARGING
STATUS OF THE BATTERIES, page 72).
The current battery status can also be queried in the menu:
Controller: Menu GLP/TOOLS -Check - Controller Info
Measuring head: Menu GLP/TOOLS - Check - info
Controller Power supply 3 x alkaline batteries (alkali manganese),
Size: Mignon, AA, AM3, LR6
Runtime: > 100 h
(approx. 1000 start-ups in normal use)
Data backup 1 x lithium battery CR2430 (accessory).
Runtime: typically 4 years
Measuring
head
Power supply Batteries: 2 x lithium battery CR2430 (accessory).
Runtime: typically 2 years
1st level (warning level) Battery LoBat !
2nd level (error level Battery empty !