10 What to do if...
10.1 Display messages
Display message Cause Remedy
already used! The selected measuring head has
already been started for a
measurement and can only be used
for a new measurement
after the current measurement is fin-
ished and the
measurement data completely
retrieved.
Select a new and free measuring
head or
- Stop start or
Release the measuring head (sec-
tion 7.6.2 R
ESET/RELEASE, page 56)
Please Set system clock! Data backup battery was changed. Set the clock (see section 7.4.3 D
ATE/
TIME, page 42).
defective! The selected measuring head is
defective. Please return it to WTW.
Select a new and free measuring
head or
- Stop start
No samples
available!
There are no finished samples stored
in the sample management.
Transfer the unfinished samples of
the sample management to “fin-
ished” status. To do this, the mea-
surement data of all measuring
heads must be called up and evalu-
ation and documentation of the
measurement data of the finished
samples must be performed.
If the measurement data of the
measuring heads are not yet
finished, no instant remedy is possi-
ble!
The memory capacity limit of the
instrument has been reached.