What to do if... OxiTop
®
Control OC 100
68 ba31114e06 06/2012
No active ! No measurement has been
started.
The measuring heads are placed
at another location (e.g. other lev-
els in the incubator, other incuba-
tors, further storage locations).
The measuring heads have no
optical contact with the controller:
Distance too great, angular
position imprecise
IR window of the measuring
head averted from the control-
ler
Shading by other objects.
Controller defective
Modification of the distances or
angle
Search at other storage locations
Check the controller
Check the measuring heads (see
section 10.3 G
ENERAL INFORMATION,
page 73)
cal test due on
18.07.98!
(sample date)
GLP is switched on
The test resource monitoring is
due within the
measuring time of measurement
for the selected
measuring head.
Date/time is not correctly set in the
controller.
Use another free measuring head
that has been checked
Start the due measuring head with
the Cal test (see section 7.5.4
CAL-
TEST, page 50)
Switch off GLP and start the mea-
suring head (if your laboratory prac-
tice permits this).
Set the date/time in the controller.
Lack of memory possi-
ble!
Memory for 11
free!
Continue
This message is used as information
in the operating mode.
There are a lot of sample data stored
in the sample management. The
memory is almost full.
It is still possible to start a sample as
a parallel sample with 11 measuring
heads (the number 11 used here
is an example, possible numbers:
1...11 ).
Problem solution, if necessary:
Erase finished samples section
7.6.1 E
RASING FINISHED SAMPLES,
page 54
Alternative: Changing the instru-
ment setting to automatic erasing of
finished samples: section 7.4.6
D
ATA STORAGE, page 43
Display message Cause Remedy