7.4.5 GLP - calibration interval (-Calinterval)
Set the time period (1 to 36 months) here. When it expires, the instrument regis-
ters that the next test resource monitoring of the measuring heads is due. After
the calibrating interval expires, measuring is blocked until the calibration is
performed or the Off setting is selected.
7.4.6 Data storage
Here, you can set whether the controller should automatically erase the oldest
finished samples if the memory is full to create space for new measurement
data (setting auto).
With the manual setting, the message "Memory lack! Erase finished sample/s!
is displayed if the memory is full (see section 10 W
HAT TO DO IF..., page 67).
GLP menu:
Preselected: -GLP
Confirm with <RUN/ENTER>.
Use <><> to select either
GLP On or Off.
Confirm with <RUN/ENTER>.
If the end of the calibrating interval set up lies within the measuring
time of a measurement to be started, this measurement cannot be
started.
GLP menu:
Select the - Calinterval menu item
with <><>
Confirm with <RUN/ENTER>.
Use <><> to set the time
period for the calibrating interval (1
- 36 months).
Delivery setting:
12 months.
Confirm with <RUN/ENTER>.
Back up your data regularly!