7.6 Maintenance
CHECK menu.
Select the - Pneumatic test
menu item with <><>.
Set the plunger of the syringe on
the OxiTop
®
PT test resource to
5 scale parts.
Tightly screw the measuring
head to be tested onto the Oxi-
Top
®
PT test resource.
(Exceeding the time causes a
return to the CHECK menu).
(Exceeding the time causes a
return to the CHECK menu).
The controller display shows the
result of the pneumatic test.
Use <RUN/ENTER> to return to
the CHECK menu.
GLP/TOOLS menu.
Select the - Maintenance menu
item with <><>.
RUN/ENTER
RUN/ENTER
RUN/ENTER