EasyManua.ls Logo

wtw OxiTop C - Page 20

wtw OxiTop C
82 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
Operating mode: BOD Standard / BOD Routine OxiTop
®
Control OC 100
20 ba31114e06 06/2012
Confirm the selected
measurement range for the
sample. The automatically
assigned sample number
(YY/MM/DD and sequential
number) is given in the header
line.
Additional information:
Type of measurement, run duration, measurement range,
final date, ID number.
Temp display => the AutoTemp function (see section 7.1
GLP/TOOLS
MAIN MENU, page 38) is switched on.
GLP display => the GLP function (see section 7
GLP/TOOLS, page 38) is switched on.
Change the ID number for the additional identification of the
sample (e.g. sampling location) as follows:
Select the
- I.D. number menu item with <><>
Confirm with <RUN/ENTER>.
Set the required ID number (setting range 001 ... 255) with
<><>.
Confirm with <RUN/ENTER>.
Use <PRINT>
to print out the entire sample information (in BOD
Routine mode only), see section 6 P
RINTING, page 36).
Confirm the start of
measurement. Contact selection:
The controller repeatedly sends the start information in the
scanner mode until successful feedback is received from the
OxiTop
®
-C measuring head. After the successful start message
from the measuring head:
Displays ! started !
From this point in time, the sample
exists within the
SAMPLE STATUS TYPE list
(sample management).
Subsequently, the request to
start the next measuring head is
made automatically.
RUN/ENTER
RUN/ENTER

Table of Contents

Related product manuals