Operating mode: BOD Standard / BOD Routine OxiTop
®
Control OC 100
30 ba31114e06 06/2012
Case 1 The sample is still running and
individual measuring heads of
the
sample are not attainable.
Case 2 The run duration of the sample
has
ended and the finished, com-
plete data of the attainable mea-
suring heads have been
transferred.
then
-Erase:
Removes missing measuring
heads from the data stock. The
completed sample is given the
“finished” status.
-Back:
Jumps back to the previous
menu.
Abort using a function key: The
stopped sample is not declared
as “finished”.
Case 3:
No measuring head of a sample
is attainable.
(In the BOD Routine operating
mode: The relevant measuring
head is not attainable).
then
-Erase:
Removes a sample from the
data stock.
-Back:
Jumps back to the previous
menu.
- Show
:
As in - Show sample, but selectively for missing measuring
heads (identification option if only the sender of the measuring
head is defective. Otherwise, see the section 10 W
HAT TO DO
IF..., page 67)
- Missing
:
Only appears when the run duration of the sample has
expired. Here you can remove the data of a missing or defec-
tive measuring head from the SAMPLE STATUS TYPE list
(sample management) of the controller. In the BOD Standard
operating mode, you can set a sample to the "finished" status
(checkmark ) if all the other measuring heads are already “fin-
ished”: Erases the missing measuring heads from the
SAMPLE STATUS TYPE list (sample management).
RUN/ENTER
RUN/ENTER