OxiTop
®
Control OC 100 GLP/TOOLS
ba31114e06 06/2012 51
Evaluation of the
Cal test
The instrument automatically
allocates the sample number (in
the header line).
The filling volume (164 ml) and
the type together with the run
time of 5 days are preset.
For information on the further handling of the sample up to the
“finished” status: see section 5.2 S
TARTING THE MEASUREMENT,
page 19.
The Cal sample appears together with the other samples in the
SAMPLE STATUS TYPE list (sample management). The BOD
type is CAL:
SAMPLE STATUS TYPE list
(sample management)
RUN/ENTER
Use <><> to select the test
sample.
Entry into the evaluation.
Display of the selected sample
as a curve with measurement
value data.
Compare measurement value
with lot test value (according to
operating manual of OxiTop
®
PM) .
Now you can set a new
calibration date for the measur-
ing head with the controller.
- Show (to find the relevant
measuring head):
Function and messages as
described (see section 5.4
SAMPLE STATUS TYPE
LIST
(
SAMPLE MANAGEMENT), page 24.
RUN/ENTER