6.12 General Specifications
Item Specifications
Twisted pair cable for the 701261, 701262, and 701265
Use a cable that is 3 m or less in length, and attach a ferrite-core
6
to the side of the cable
closest to the DL850E/DL850EV by passing the cable twice through the core.
Twisted pair cable for the 720220 and 720221(701953)
Use a cable that is 3 m or less in length, and attach a ferrite-core
8
to the side of the cable
closest to the DL850E/DL850EV by passing the cable twice through the core.
Measurement lead 758933 (for the 701267)
Bundle the H and L measurement leads. Attach a ferrite core
6
to the DL850E/DL850EV side
of the bundled leads.
D-sub 9-pin shielded cable (for the 720240 and 720241)
Use a cable that is 3 m or less in length, and attach a ferrite core
6
to the side of the cable
closest to the DL850E/DL850EV.
Bridgeheads 701955 and 701956 (for the 701270)
Use a cable that is 5 m or less in length, and attach a ferrite core
6
to the side of the cable
closest to the DL850E/DL850EV.
Bridgeheads 701957 and 701958 (for the 701271)
Use a cable that is 5 m or less in length, and attach a ferrite core
6
to the side of the cable
closest to the DL850E/DL850EV.
Logic probe input cables 700986, 700987, 702911, and 702912 (for the 720230)
Attach a ferrite core
6
to the side of the cable closest to the DL850E/DL850EV.
Immunity Compliant standards
EN61326-1 Table 2 (for use in industrial locations), EN61326-2-1
Approved (DL850E, DL850EV, 701250, 701251, 701255, 701267, 701261, 701262, 701265,
701270, 701271, 701275, 701281, 720210, 720211, 720220, 720221, 720230, 720240, 720241,
720243, 720254, 700986, 700987, 701953, 701955, 701956, 701957, 701958, 702911, 702912)
Influence in the immunity environment (criteria A)
Noise increase
701250: ≤±20 mV (1:1 input, 5 mV/div conversion)
701251: ≤±3 mV (1:1 input, 1 mV/div conversion)
701255: ≤±25 mV (1:1 input, 5 mV/div conversion)
701267: ≤±30 mV (1:1 input, 20 mV/div conversion)
701261: ≤±3 mV (5 mV/div range conversion)
701262: ≤±3 mV (5 mV/div range conversion)
701265: ≤±0.05 mV (0.1 mV/div)
701270: ≤±100 μSTR (±500 μV range, gauge factor = 2)
701271: ≤±100 μSTR (±500 μV range, gauge factor = 2)
701275: ≤±6 mV (1:1 input, 5 mV/div range conversion)
701281: ≤±0.01 Hz (frequency, 0.1 Hz/div conversion)
720210: ≤±50 mV (1:1 input, 10 mV/div range conversion)
720211: ≤±50 mV (1:1 input, 10 mV/div range conversion)
720220: ≤±20 mV (0.2 V/div range conversion)
720221: ≤±2 mV (1 mV/div range; data update period = 100 ms when you are using the
701953)
720230: No bit errors
720240: No erros
720241: No erros
720243: No erros
720254: ≤±15 mV (1:1 input, 10 mV/div conversion)
Test conditions
701250: 10 MS/s, envelope mode, 50 mV/div, no input filter, with the tip of the probe (700929
(10:1)) shorted
701251: 1 MS/s, envelope mode, 10 mV/div, no input filter, with the tip of the probe (700929
(10:1)) shorted
701255: 10 MS/s, envelope mode, 50 mV/div, no input filter, with the tip of the probe (701940
(10:1)) shorted
701267: 100 kS/s, envelope mode, 0.2 V/div, no input filter, with the end of the cable shorted
701261: 100 kS/s, envelope mode, 5 mV/div, no input filter, with the end of the cable shorted
701262: 100 kS/s, envelope mode, 5 mV/div, no input filter, with the end of the cable shorted
701265: 500 S/s, envelope mode, 0.1 mV/div, no input filter, with the end of the cable shorted
701270: 100 kS/s, envelope mode, 500 μSTR, gauge factor: 2.0, no input filter
701955 bridge voltage: 2 V
701956 bridge voltage: 10 V
701271: 100 kS/s, envelope mode, 500 μSTR, gauge factor: 2.0, no input filter
701957 bridge voltage: 2 V
701958 bridge voltage: 10 V
701275: 100 kS/s, envelope mode, 50 mV/div, no input filter, with the tip of the probe (700929
(10:1)) shorted
701281: 1 MS/s, envelope mode, Frequency, no input filter, with the tip of the probe shorted