13-2
IM WT3001E-51EN
•  Limitations on the Function and Measurement Function
  In cycle-by-cycle measurement mode, internal computation is processed differently 
from other measurement modes. Therefore, some functions such as waveform 
and bar graph displays on the screen and storage function cannot be used in this 
measurement mode. In addition, some measurement functions such as integration 
and efficiency computation cannot be carried out. For a list of functions with limitations 
and measurement functions that cannot be measured, see appendices 10 and 11 in 
the User’s Manual IM WT3001E-01EN.
Note
In cycle-by-cycle measurement, the equation for apparent power and reactive power is fixed to 
TYPE 2. If TYPE 1 or TYPE 3 is selected and you select cycle-by-cycle measurement mode, 
the equation switches to TYPE 2. For details on the type of equations, see section 2.5 and 5.9 
in the User’s Manual IM WT3001E-01EN.
•  Data Update Interval
  The data update interval setting is invalid in cycle-by-cycle measurement mode. The 
display data is updated when the measurement of the specified number of cycles 
is completed. The display data is not updated until then. For example, if you are 
measuring a 50-Hz (20-ms) input signal for 1000 cycles, it takes at least 20 s from 
the start of the measurement until the measurement is completed and the measured 
values are displayed.
•  Saving the Cycle-by-Cycle Measurement Data
  To save the cycle-by-cycle measurement data on this instrument, select Numeric for 
the item to be saved. For details, see section 9.3 in the User’s Manual IM WT3001E-
01EN.
13.1  Cycle-by-Cycle Measurement Function