13-3
IM WT3001E-51EN
Cycle-by-Cycle Measurement
13
• Limitations on the Setting Changes and Execution of Operations in
Cycle-by-Cycle Measurement Mode
There are some functions whose settings cannot be changed in cycle-by-cycle
measurement mode as shown below.
Setup Item Cycle-by-Cycle Measurement Status
Reset Start Complete
Cycle-by-cycle measurement
Synchronization source and trigger Yes View View
Cycle count and timeout Yes View View
Measurement start Yes No No
Reset – Yes Yes
Displayed items Yes Yes Yes
Displayed cycle number Yes Yes Yes
Measurement mode Yes No No
Wiring and compensation
Wiring system Yes View View
Independent setting of elements Yes View View
Efficiency and compensation View View View
Voltage/current range Yes No No
Auto range No No No
Sensor transformation ratio Yes View View
Crest factor Ye s View View
Voltage/current mode Yes No No
Scaling Yes View View
Filter Yes View View
Averaging No No No
Synchronization source No No No
Measurement and computation ([MEASURE]) No No No
Frequency measurement No No No
Cursor – – –
NULL No No No
Motor evaluation Yes View View
Harmonic measurement No No No
Displayed items [U/I/P][S/Q/λ/φ][FU/FI/η] Yes Yes Yes
[WP/q/TIME] [USER] – – –
[ELEMENT] Yes Yes Yes
Data update interval No No No
Hold Yes Yes Yes
Single measurement (display update) Yes Ye s Yes
Zero-level compensation Yes No No
Integration No No No
File operation Yes Yes Yes
Set or execute the screen image storage Yes Yes Ye s
Set or execute the storage operation No No No
Set or execute the print operation Yes Yes Yes
Miscellaneous ([MISC]) Yes Yes Yes
Help Yes Yes Yes
Yes: The settings can be changed and the function can be executed.
No: The settings can be changed and the function cannot be executed.
View: The settings can be viewed but cannot be changed, and the function cannot be executed.
–: Does not function.
13.1 Cycle-by-Cycle Measurement Function