Carl Zeiss Requirements on Microscope Stand Cell Observer SD
40 423638-7044-001 M60-2-0021 e 01/2009
3.2 Requirements on Microscope Stand
The Cell Observer SD has been formally approved for combined operation with these types of microscope
stand:
Axio Observer.Z1 (431007-0000-000 or 431007-9901-000)
The manufacturer does not assume liability for any use that is unrelated to the
intended use.
Combining the Cell Observer SD with an Axio Observer stand of an ordering code number that
is different from those quoted herein will be deemed as non-conforming use.
In particular, no warranty can be assumed by Carl Zeiss MicroImaging GmbH regarding laser
safety issues.
Because the system includes highly magnifying objective lenses, it is recommended to work
with a vibration-prevention device. Place the microscope on a vibration-dampened table or an
anti-vibration plate, in order to ensure absence of mechanical vibrations.
For long-time experiments it is necessary to maintain a stable ambient temperature, in order to
prevent artefacts due to thermal extension.
Further required components for connection to the microscope:
parts for CSU-X1-to-Axio Observer matching,
basic laser safety kit,
CSU-X1 laser safety kit.