12-12 Nightingale PPM3 User's Guide April 2019
Managing Alarms
NBP Conditions:
Condition
Display
Value
Alarm
Grade
Message
Annunciation
Type
NBPs within limits
<NUM>
None
None
None
NBPs < LL
<NUM>
Medium
NBPs < LL
One Time
NBPs > UL
<NUM>
Medium
NBPs > UL
One Time
NBPm within limits
<NUM>
None
None
None
NBPm < LL
<NUM>
Medium
NBPm < LL
One Time
NBPm > UL
<NUM>
Medium
NBPm > UL
One Time
NBPd within limits
<NUM>
None
None
None
NBPd < LL
<NUM>
Medium
NBPd < LL
One Time
NBPd > UL
<NUM>
Medium
NBPd > UL
One Time
NBP Pulsations Too Small
---
Low
NBP weak signal
One Time
NBP Too Much Motion
---
Low
NBP artifact
One Time
NBP Leaky Cuff or Hose
---
Low
NBP cuff leak
One Time
NBP Pinched Hose
---
Medium
NBP blocked hose -- check
patient
One Time
NBP Measurement Time-out
(2¼ minutes)
---
Low
NBP measurement time
exceeded
One Time
NBP Pump or Valve Failure or
NBP Safety Timer Expired or
Other H/W-related problem
---
Medium
NBP problem detected
One Time
NBP Bad Profile Shape
---
Low
NBP cannot measure
One Time
CO
2
Conditions:
Condition
Display
Value
Alarm
Grade
Message
Annunciation
Type
ETCO
2
within limits
<Number>
None
None
Persistent
ETCO
2
< [lower limit]
<Number >
Medium
ETCO
2
< LL
Persistent
ETCO
2
> [upper limit]
<Number >
Medium
ETCO
2
> UL
Persistent
FICO
2
within limits
<Number>
None
None
Persistent
FICO
2
> [upper limit]
<Number >
Medium
FICO
2
> UL
Persistent
RRc
within limits
<Number>
None
None
Persistent
RRc < [lower limit]
<Number >
Medium
RR < LL
Persistent
RRc > [upper limit]
<Number >
Medium
RR > UL
Persistent
ETCO
2
out of range (high)
<Number >
Medium
CO
2
out of range (high)
Persistent