Service Kit Installation Procedures
4–58 www.zoll.com 906-0731-04-01 Rev. C
2 Loosen and remove the two 10-32 Keps nuts and
the four 6-32 x 2 screws.
Save these screws.
Remove the back case by lifting from the
ventilator.
3 Remove the two 4-40 x 1/4 screws on the dovetail
mounting bracket.
Save these screws.
Disconnect the ribbon cable on the PIM board by
simultaneously applying pressure on the two
ejector latches.
4 Flip the ventilator over and remove the front case
assembly by lifting it straight up away from the
ventilator module.
5 Using a sharp knife, carefully cut the RTV sealant
around the SpO
2
cable.
6 Loosen and remove the two M2.5 x 5mm screws
holding the SpO
2
cable to the front case.
Step Action