Ventilator & Calibration Troubleshooting
906-0731-04-01 Rev. C ZOLL Ventilator Service Manual 6–55
3450
Battery Discharge Fault
Alarm triggers when the battery temperature reaches 70 °C (158 °F) which is
5 °C from its maximum operating temperature and external power is
connected.
4 Year Maintenance Kit
3451
Battery Discharge Fault
Alarm triggers when the battery temperature reaches ≥75 °C (167 °F) and
external power is connected.
4 Year Maintenance Kit
3452
Battery Charging Fault
Alarm triggers when the battery temperature is > 45 °C
(122 °F).
4 Year Maintenance Kit
3453
Battery Charging Fault
Alarm triggers when the battery temperature is ≤ 0 °C (32 °F).
4 Year Maintenance Kit
3455
Battery Fault
Alarm triggers when the device is not able to communicate with the
internal battery and external power is connected.
Power Interface Module
Service Kit
4 Year Maintenance Kit
3470
Self Check Fault
Alarm triggers when the device is no longer able to communicate with the
Power Interface Module (PIM).
Check cable from PIM board
to CPU
Power Interface Module
Service Kit
3480
Self Check Fault
Alarm triggers when the device software detects that it has not been
calibrated with the SPM that is inside the device.
Recalibrate using the RCS
Service Code Associated Alarm Name/Description Service Kit/Resolution