Maintenance Schedule & Procedures
3–6 www.zoll.com 906-0731-04-01 Rev. C
7
Note: This test is only required
in the case of a reported
complaint or identified failure.
TCE 7.X Incoming System Tests
RCS 8.X (System Tests)
TCE 7.X - From the Sequence list, select & run “Incoming System Tests”
RCS 8.X - From the Service Actions list, select and run “System Tests”
If the test fails, check your set up and re-try. Review Troubleshooting guide to
address failures. Use the Discrete Test drop down to confirm that the failed result
has been addressed Prior to running the calibration and outgoing system test.
Do NOT retry more than 3 times. Call for support.
Note: Completion of the Incoming System Test Sequence is not required. Verify
failures have been corrected using the discrete tests.
8
Device Service or PM Kit
Installation
Service the device to address issues identified during testing.
Install 1 Year or 4 Year PM Kit per instructions.
9
TCE 7.X - Complete Calibration
RCS 8.X - Calibration
TCE 7.X - From the Sequence list, Select & Run “Complete Calibration”
Note: Complete Calibration must complete without any failures to apply the
calibration to the device.
RCS 8.X- From the Service Actions list, Select & Run “Calibration”
Note: Calibration must complete without any failures to apply the calibration to
the device.
10
TCE 7.X - Outgoing System Test
RCS 8.X (System Tests)
TCE 7.X - From the Sequence list, select & run “Outgoing System Tests”
RCS 8.X - From the Service Actions list, select and run “System Tests”
Do NOT retry more than 3 times. Call for support.
Note: Outgoing System Tests / System Tests must be completed without any
failures to update the PM due date and generate the customer certificate.
11 Erase Forensic Memory From the Sequence or Service Actions list, Select & Run “Erase Forensic Memory”
12 Final Configuration
TCE 7.X - From the Sequence List, select and run “Final Configuration”, select the
appropriate configuration file.
RCS 8.X - Final Configuration is only required only when a change in the current
configuration is necessary.
Note: Do NOT change the original configuration without prior authorization.
Contact technical support to change configuration or for authorization.
13
Test Records & Calibration
Reports
Upon completion of the testing and calibration sequences, ensure calibration
reports are printed, signed and filed.
Table 3
Step Description & Notes