Maintenance Schedule & Procedures
906-0731-04-01 Rev. C ZOLL Ventilator Service Manual 3–7
Stockpile Warm-Up Procedure
It is recommended for ventilators not in regular continued use to use the following settings to exercise the electro-
mechanical components prior to connecting the device to the RCS system. Failure to complete this warm-up
procedure may result in testing or calibration failures of the device.
Step Instructions
1. Connect 55 PSI O2 to the ventilator.
2.
Connect the patient circuit to the vent, the circuit’s patient connection end to resistor R#2
then add the test lung.
3.
On the vent set Mode: AC(V), Breath Rate= 20 BPM, Vt= 450ml, PIP High alarm=50cmH2O,
FIO2=60% run for 5 minutes.
4. Set FIO2 to 100%. Run for 5 minutes.
End of Procedure