42/24-10 EN Rev. 9 Chapter 10: Status Messages, Troubleshooting 10-15
Status Messages, continued
No. Text O E M F Reaction /Remark
509 Automatic calibration started. For information
510 Automatic calibration ended. For information
511 Automatic calibration
externally interrupted.
For information
512 Automatic calibration in
progress.
x For information, not logged
513 System bus communication
defect during automatic
calibration.
514 External calibration started. For information
515 External calibration ended. For information
516 External calibration in
progress.
x For information, not logged
517 Device being serviced. x For information, e.g. during manual calibration, not logged
518 The calibration could not be
done, because the value is not
stable.
519 Preamplifier overflow error:
Calibration could not be
performed because of
preamplifier override.
520 Initial zero calibration started. For information
521 Initial zero calibration ended. For information
522 Initial zero calibration
interrupted.
For information
523 Initial zero calibration
incomplete. System bus
communication defect during
calibration.
For information
524 Initial zero calibration in
progress.
x For information, not logged
525 Linearization impossible:
Linearization did not produce
a valid result. Measurement
value possibly inaccurate.
Check center point gas.
See message text
526 Linearization impossible:
Linearization could not be
performed Ie the characteristic
is linear.
See message text
527 Initial calibration for
component:
For information
528 Autocalibration not started Ie
manual calibration was
running.
For information
Continued on next page