Parameterization
OI/FEX300/FEX500-EN FEX300, FEX500 105
Menu / Parameter Value range Description
Input/Output / Logic Setup
DO1 Signal
No Function, F/R Signal,
Alarm Signal, Dual Range 1,
Batch end contact 1
The menu is displayed only if in the " DO1 / DO2
Function " menu the Logic / Logic function is set. This
menu is not displayed in the factory default setup.
• F/R Signal:
The digital output signals the flow direction.
• Alarm-Signal:
The digital output functions as alarm output. The
alarm type is set in the DO1 Alarm Config menu.
• Dual Range:
The digital output is activated when flow range 2
(Qmax 2) is selected.
• Batch end contact:
The digital output is activated when the configured
fill quantity is reached.
Default setting: F/R Signal.
DO1 Alarm Config
Select the "DO1 Alarm Config" submenu.
The men
u is displayed only if in the " DO1 Signal "
menu the Alarm Signal function is set.
DO1 Action
Normally Open, Normally
Closed
Select the switching behavior for the digital output.
Default setting: Normally Open.
DO2 Signal
No Function, F/R Signal,
Alarm Signal, Dual Range 1,
Batch end contact 1
Refer to the description "DO1 Signal".
DO2 Alarm Config
Select the "DO2 Alarm Config" submenu.
The men
u is displayed only if in the "DO2 Signal"
menu the "Alarm Signal" function is set.
DO2 Action
Normally Open, Normally
Closed
Refer to the description "DO1 Action".
italics = Parameter can only be viewed at the "Advanced" password level.
1) Parameter / menu only available for FEP500 / FEH500.