Parameterization
118 FEX300, FEX500 OI/FEX300/FEX500-EN
Menu / Parameter Value range Description
Diagnostics / Diagnosis Control / Gas Bubble Detector
1)
Gas Bubble Detector On / Off Activate the "Gas Bubble Detector" function.
Default setting: Off
Important (Notice)
The gas bubble detector can be used in the
nominal diameter range DN10 ... 300.
For further information read and observe chapter
“Extended diagnostic functions”.
Adjust Gas Bubble
The Gas Bubble Detector must be adjusted according
to the conditions on site. Start automatic adjustment of
the Gas Bubble Detector function.
Important (Notice)
Prior to starting the adjustment, make sure that:
• There is no flow through the flowmeter sensor
(close all valves, shut-off devices, etc.)
• The flowmeter sensor is completely filled with
the fluid to be measured and free from gas
bubbles.
Threshold Gas Bubble
Set the switching threshold. If the current value
exceeds the defined switching threshold, a message
appears on the display and an alarm is output via the
digital output, if configured accordingly.
Gas Bubble Value Display only Display the current gas bubble value.
Diagnostics / Diagnosis Control / Coating Detector
1)
Coating Detector On / Off Activate the "Electrode Deposit Detector" function.
Default setting: Off
Important (Notice)
The electrode deposit detector can be used in the
nominal diameter range DN10 ... 300.
For further information read and observe chapter
“Extended diagnostic functions”.
One shot Coating
The electrode deposits are measured cyclically at
defined intervals. Here, a current measurement can
be started.
Coating Value QE1 Display only Current deposit value for electrode 1
Coating Value QE2 Display only Current deposit value for electrode 1
italics = Parameter can only be viewed at the "Advanced" password level.
1) Parameter / menu only available for FEP500 / FEH500.