EasyManua.ls Logo

ABB FSM4000 - Current Output Settings

ABB FSM4000
136 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
FSM4000 ELECTROMAGNETIC FLOWMETER | OI/FSM4000-EN REV. E 75
Submenu / parameter Value range Remark
Totalizer → F
Numerical entry
0 ... 9999999
Display or input the forward totalizer value.
0.0 kg/h
Overflow ← R
Read only Overflow totalizer max. 65535. 1 overflow = pulse totalizer
9,999,999 units. Resets the display status and increments the
overflow counter by one.
0
Totalizer ← R Numerical entry
0 ... 9999999
Display or input the reverse flow totalizer value.
0
Reset totalizer Select ENTER to reset the forward/reverse flow totalizer and
overflow counters.
Overflow Diff. Read only Display the overflow for the diff. totalizer.
0
Totalizer Diff. Numerical entry
0 ... 9999999
Display or input the diff. totalizer value.
0.0 kg/h
Submenu This menu does not appear for the PROFIBUS PA or
FOUNDATION fieldbus.
Current output
Current output
0 - 20 mA, 4 - 20 mA
0 - 10 mA, 2 - 10 mA
Select the current output range.
For HART protocol always 4 - 20 mA.
0 - 20 mA
Iout for alarm Low Alarm
High alarm
Behavior of the current output during an alarm. Low-Alarm is
available only for current output of 2 - 10 mA and 4 - 20 mA. If
error 0 (empty tube) occurs, the current output is set to the value
programmed for the parameter ‘lout during empty tube‘ in the
submenu ‘Detector empty tube‘. If error 3 (flow rate > 103 %)
occurs, ‘High Alarm’ is always signaled.
Low Alarm
Iout low alarm
Numerical entry Low-Alarm is available only for current output of 2 -10 mA or 4 -
20 mA. Set the current output value during Low Alarm condition.
No selection available if High Alarm is selected for the current
output selection for lout during alarm and if the transmitter
version is standard and the current output range is
0 - 20 mA
0 - 10 mA
.
3.60 mA

Table of Contents

Other manuals for ABB FSM4000

Related product manuals