84 FSM4000 ELECTROMAGNETIC FLOWMETER | OI/FSM4000-EN REV. E
8 Operation
Parameter descriptions
Submenu / parameter Value range Remark
Simulation Mode
Off
A
On/Off function for Simulation mode. Before operation with
flow, a flowrate can be simulated. The output values
correspond to the simulated flowrate entered. The
information ** Q Simulation** is displayed in the lower line of
the display. After completing the simulation, change the
parameter in the Simulation Mode menu to ‘Off’.
Off
Simulation Value
Numerical entry
-130 % - +130 %
Select the simulation value for the display and outputs.
Current and pulse output follow the flowrate value.
0.0 %
Submenu Menu available only if the required hardware is available.
Diagnosis
Diagnosis
Menu is displayed only with ‘Manual’ diagnosis interval.
Manual
Start? Yes, exit via C/CE Manually start manual diagnostic. The value is displayed after
a delay of 30 s.
Yes → Enter
Submenu
Diag. Interval
Cycle Time
Manual; 10 s; 60 s; 10
min; 60 min; 6 h; 12 h;
24 h; 7 d
Select the time interval between the individual diagnostic
measurements. For the values for Coil Resistance, Coil
Temperature, Insulation Resistance, and Magnetic Linearity,
however, there are minimum measurement intervals for
technical reasons related to the measurement:
60 seconds for Coil Resistance and Coil Temperature
5 minutes for Magnetic Linearity
10 minutes for measurement of Insulation Resistance
These times are not undershot independently of the
configured diagnostic cyclic.