7-8
Questions? Refer to Page 11-22 For Technical Support Instructions.
Chapter 7: Linearity Test
What You See on the Display What You Do Comments
100 mg/dL
15. You can select one of the following
options:
Press 1–New Level.
Press 2–Same Level.
Press the Menu button to return to the
Menu Mode.
Press On/Off to turn off meter.
Remove the test strip from the meter
and discard it when finished testing.
Follow your facility’s biohazard disposal
policy.
The display shows:
Level (Test level 1 to 5)
Test (Replicate 1 to 4 for that level)
Comment Code (Number in box)
Test Type (Glucose)
Date and Time of test
Operator ID (o: prompt)
Note that test date and time alternates with
Operator ID.
If 2 is selected, the Scan or Enter Strip
Lot screen will appear. Return to step 8 to
continue linearity testing with the Same
Level.
If 1 is selected, the Select Level screen will
appear. Return to step 7 to perform linearity
tests for a New Level.