EasyManua.ls Logo

Abbott FreeStyle Precision Pro - Page 105

Abbott FreeStyle Precision Pro
186 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
7-7
Questions? Refer to Page 11-22 For Technical Support Instructions.
Chapter 7: Linearity Test
What You See on the Display What You Do Comments
13. Wait for the meter to analyse the sample
and display the test result.
When using FreeStyle Precision Pro
test strips the countdown will begin at 5
seconds. When using Precision Xceed
Pro test strips the countdown will begin
at 20 seconds.
The meter counts down then displays the
test result. If an error is detected during
the assay, the assay is terminated and no
result is displayed. Once the assay begins,
all key presses are ignored until the assay is
completed.
100 mg/dL
14. Note the result. If required, scan or
manually enter the comment code and
press Enter.
If there is no prompt to enter a
comment code, skip down to step 15.
If the comment code option is enabled in
the control test configuration through the
data management system, the meter will
be enabled to scan or enter a 1 or 2 digit
comment code.

Table of Contents