4-10
Questions? Refer to Page 11-22 For Technical Support Instructions.
What You See on the Display What You Do Comments
11. You may select one of the following:
Press 1 – Next Level.
Press 2 – Repeat Test.
Press Menu to return to the
Menu Mode menu.
Press On/Off to turn off the meter.
Remove the test strip from the meter
and discard it when finished testing.
Follow your facility’s waste disposal
policy.
The display shows:
Level (Low, Mid, High)
Comment Code (Number in box)
Test Type (Glucose)
Date and Time of test
Operator ID (o: prompt)
Note that test date and time alternates with
Operator ID.
If 1 is selected, the Scan or Enter Control
Lot screen will appear. Return to step 4 to
perform the Next Level control test.
If 2 is selected, the Scan or Enter Strip
Lot screen will appear. Return to step 5 to
Repeat Test at the same level.
Note: If ketone testing is disabled, then following a complete set of Glucose controls, 1-Next Level becomes
1-Exit.
If ketone testing is enabled and Glucose QC is run first, then following a complete set of Glucose
controls, 1-Next Level becomes 1-QC Ketone.
If ketone testing is enabled and Ketone QC is run first, then following a complete set of ketone controls,
1-Next Level becomes 1-QC Glucose.
Chapter 4: Control Test