6-5
Questions? Refer to Page 11-22 For Technical Support Instructions.
What You See on the Display What You Do Comments
9. Ensure proficiency survey specimens
are at room temperature. Mix each
specimen well according to instructions
from the survey provider. Wipe away
any specimen on the tip of the vial
before squeezing the vial and applying
a drop to the test strip target area. After
applying the sample, recap the vial
tightly.
When sufficient sample has been applied,
the meter beeps, displays Sample
Accepted and automatically starts the test.
For a detailed description of the test strip
target area, refer to the Test Strips section of
Chapter 1.
10. Wait for the meter to analyse the sample
and display the test result.
When using FreeStyle Precision Pro
test strips the countdown will begin at 5
seconds. When using Precision Xceed
Pro test strips the countdown will begin
at 20 seconds.
The meter counts down then displays the
test result. If an error is detected during
the assay, the assay is terminated and no
result is displayed. Once the assay begins,
all key presses are ignored until the assay is
completed.