Art: 730077-00B Rev. Date: 14-Mar-12
18
Note 3.1: If this message appears, users can press the listed number on the handheld keypad
to display which cartridges are currently inactive.
4. “i-STAT QC Past Due – Cartridge Testing Disabled”: This message will appear if the liquid
QC requirements for a schedule were not completed within the grace period, and if the
handheld is configured for “non-information-first” or “information-first with cartridge lot
number disabled”. In these scenarios, if any cartridge type is locked out, all cartridge
types are disabled.
Note: When the handheld is customized for multiple liquid QC schedules, there is a hierarchy
which determines which of the four messages above will appear on the Test Menu display. The
messages will appear in this order:
i-STAT QC Past Due – Cartridge Testing Disabled
QC Past Due – Not All Cartridges Are Active For Testing
i-STAT QC Due – Complete Before (Date)
Next i-STAT QC (Date)
Quality Tests Menu: When testing liquid QC material and the user has selected 1-Control in the Quality
Tests Menu on the handheld, a new “Select QC Event” screen will then appear on the display.

Table of Contents

Other manuals for Abbott i-STAT 1

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Abbott i-STAT 1 and is the answer not in the manual?

Abbott i-STAT 1 Specifications

General IconGeneral
BrandAbbott
Modeli-STAT 1
CategoryMeasuring Instruments
LanguageEnglish

Related product manuals