TECHNICAL BULLETIN
i-STAT
®
ANALYZER CODED MESSAGES
From the time it powers up until the time it powers down, the i-STAT
®
Analyzer performs numerous quality
checks. The failure of any quality check causes the analyzer to halt the test cycle and display a “cause”, an
“action” message, and a code.
The Cause Message:
This message describes the likely cause of the failed quality check. For example, when an overfilled
cartridge is detected, the analyzer will display “Sample Positioned Beyond Fill Mark”.
The Action Message:
This message indicates the appropriate action. For example, if it is likely the quality check will fail again the
next time the analyzer is used, the instruction “Use Electronic Simulator” will be displayed. If the problem is
related to an operator or cartridge, the instruction “Use Another Cartridge” will be displayed.
The Cause Code:
This is a numeric code associated with the failed quality check. Since multiple codes can be associated
with a single cause message, this is essential information when contacting i-STAT Technical Services or your
local support organization for further assistance. The codes are stored in the analyzer’s memory along with
other test records and are transmitted to the Central Data Station. The code list can be viewed and printed.
Codes 1-15 and 95 usually indicate a condition related to the environment or the state of the analyzer.
These conditions are usually benign and go away after the next cartridge or Electronic Simulator is
inserted, or after the offending condition is corrected.
Code
Number
Cause/Action
Message on Display
Explanation
1
Dead Batteries /
Replace Batteries
There is insufficient battery power to complete the testing
cycle. Replace the disposable lithium batteries in the analyzer
or recharge the rechargeable batteries.
If you are experiencing this code frequently and use disposable
batteries with the i-STAT 1 analyzer, you may want to consider
the rechargeable battery system available with the i-STAT 1
Analyzer.

Table of Contents

Other manuals for Abbott i-STAT 1

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Abbott i-STAT 1 and is the answer not in the manual?

Abbott i-STAT 1 Specifications

General IconGeneral
BrandAbbott
Modeli-STAT 1
CategoryMeasuring Instruments
LanguageEnglish

Related product manuals