6 - 2
Art: 714368-00K Rev. Date: 02-Aug-12
Specifications
Power Supply
Specification Downloader and
Downloader/Recharger
Input 100 - 240V~
47 - 63 Hz
.9 - .5A
Output 12V
3A max
* Recharge feature cannot be used in this configuration.
Specification Downloader Downloader/Recharger
Size 5.25in (13.3cm) Wide
6.75in (17.2cm) Long
2.13in (5.4cm) High
4.12in (10.4cm) Wide
10.25in (26.cm) Long
5.00in (12.7cm) High
Weight 0.6 lbs (0.27kg) 1.2 lbs (0.55kg)
Power AC-DC power adapter
or PC/Downloader adapter.
Input 12V
AC-DC power adapter
or PC/Downloader adapter.*
Input 12V
Operating
Temperature
0 to 40˚C
32 to 104˚F
0 to 40˚C
32 to 104˚F
Storage Temperature -20 to 50˚C
-4 to 122˚F
-20 to 50˚C
-4 to 122˚F
Pollution Degree (Allowable
ambient pollution level)
2 2
Installation Category (Allowable
overvoltage specification)
2 2
Communication To Central Data
Station and other equipment
Serial (RS232), or
Ethernet
Serial (RS232), or
Ethernet
Communication Link To and
From Analyzer
Infrared Transceiver Infrared Transceiver
Indicator LEDs
Power
Proximity
Charge
Green
Red
NA
NA
Blue
Red/Green
Configuration By host computer By host computer
Running Cartridges
in an Analyzer
Docked in a
Downloader/
Recharger
All i-STAT Cartridges may be run in Handhelds that are docked in a Downloader/
Recharger.

Table of Contents

Other manuals for Abbott i-STAT 1

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Abbott i-STAT 1 and is the answer not in the manual?

Abbott i-STAT 1 Specifications

General IconGeneral
BrandAbbott
Modeli-STAT 1
CategoryMeasuring Instruments
LanguageEnglish

Related product manuals