9-2 Art: 714371-00E Rev. Date: 06-Aug-12
ANALYZER CUSTOMIZATION OPTIONS AND DEFAULT SETTINGS
Option Description Default Comments
LANGUAGE WINDOW Language for text: English, Japanese, German, Italian, Dutch,
Spanish, French, Swedish, Portuguese, Danish, and Finnish
English
UNIT SET WINDOW Reporting units for results. Selected from predefined sets or by
analyte.
Unit Set 00 See table below with 17 predefined unit sets. Unit Set
99 allows the name and units for each test to be defined
individually.
Note: Reference Ranges, Action Ranges, and Custom
Reportable Ranges (if applicable) in the Preferences Window
must be changed when changing units.
i-STAT 1 ANALYZER
AND PHILIPS BAM
CLEW WINDOWS
Standardization data. All non-expired versions listed. The CLEW software has an expiration date. If an expired
CLEW remains in a customization profile, a warning will
be displayed.
i-STAT 1 SOFTWARE
WINDOW
JAMS functionality data. Users can remotely request a JAMS update for an i-STAT
1 Analyzer from the CDS. See the "Network Options for
Updating the i-STAT 1 Handheld" Technical Bulletin in the
"Care and Software Update" tabbed section of the i-STAT 1
System Manual for full details..
PREFERENCES
WINDOW
Options and default settings are listed under six headings:
Instrument, ID Entry, Test, Cartridge QC, Results, and Analyte
Enable.
STATNotes
Feature allows users to customize the Chart Page on their i-STAT
1 Analyzers in order to capture user-defined information such as
ventilator settings.
CHART0 See the "i-STAT 1 Analyzer Chart Page Customization"
Technical Bulletin for full details.
USE eVAS
This Feature can automatically determine if the results of a liquid QC
test run on an i-STAT cartridge are within APOC's published quality
control ranges.
Not enabled See the "Liquid Quality Control Pass/Fail Customization on
the i-STAT 1 Handheld" Technical Bulletin for full details.
USE OPERATOR LIST 4000 operator IDs can be stored in the analyzer along with
certification start and end dates for cartridge testing.
Not enabled (no
information stored)
Operator lists are created in the Operator Workspace on the
Central Data Station. This check box cannot be enabled if
the Operator List is empty in the Operator Workspace for all
Departments (other than the one labeled "Unassigned").