11-2 Art: 714373-00F Rev. Date: 17-Jan-11
STEP ACTION
1 Place the cartridge on a flat surface or hold it in a horizontal
position.
2 Direct the tip of the syringe, capillary tube
or dispenser into the sample well.
3 Dispense sample slowly and steadily until
it reaches the fill mark indicated on the
cartridge label. Leave some sample in the
sample well.
4 Fold the snap closure over the
sample well.
5 Press the rounded end of the
closure until it snaps into place.
Slightly lift finger or thumb
and ensure that the cartridge
is closed before completely
removing finger or thumb
from the closure.
FILLING AND SEALING CARTRIDGE USING TRANSFER DEVICE
Procedure
FILLING AND SEALING PT/INR (PROTHROMBIN TIME) CARTRIDGES USING DIRECT
FINGERSTICK SAMPLING
STEP ACTION
1. Remove cartridge from foil pouch and place the cartridge on a flat
surface.
2. Prepare lancet device and set aside until needed.
3. Clean and prepare the finger to be sampled using a 70% aqueous
solution of isopropanol (70% v/v). Allow the finger to dry
thoroughly before sampling. When disinfecting fingerstick skin
puncture sites, swabs or solutions containing substances other than
isopropanol (e.g. Chlorhexidine Gluconate) are not recommended.
Refer to the “Limitations of the i-STAT
PT/INR Test” section in the PT/INR
Cartridge and Test Information Sheet
for more information.
4. Prick the bottom side of the fingertip
with the lancet device.
5. Gently squeeze the finger, developing
a hanging drop of blood and perform
the test with the first sample of blood.
Avoid strong repetitive pressure (“milking”) as it may cause hemolysis or
tissue fluid contamination of the specimen.
Caution
Do not hold cartridge between the fingers
if using a syringe with needle to fill.
PT/INR