EasyManuals Logo

Abbott i-STAT 1 System Manual

Abbott i-STAT 1
658 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Page #245 background imageLoading...
Page #245 background image
Art: 714377-00K Rev. Date: 31-Jan-12 15-9
For example: the microhematocrit method mean for the mid level
sample is 36%PCV. The i-STAT method mean is 34%PCV. 34 x
1.0425 = 35.445. Acceptable range for the adjusted i-STAT mean: 33
- 39%PCV.
Note: If your analyzers are customized for K
2
EDTA/Heparin/None, the above
calculation is unneccessary.
Notes on the
Procedure
1. If a higher hematocrit value is needed in tube 1 or 3, packed cells can
be obtained by centrifuging the whole blood retained from tube 1 in
step 3. If a lower hematocrit value is needed, add plasma retained in
step 5.
2. The highest hematocrit that should be tested on the i-STAT System is
75%. Whole blood samples with hematocrit values greater than 75%
will be flagged as >75. The lowest hematocrit that should be tested
on the i-STAT System is 10%. Whole blood samples with hematocrit
values less than 10% will be flagged as <10.
Using Another
Comparative Method
Methods other than the reference microhematocrit procedure may be used to
verify calibration and reportable range of the i-STAT hematocrit. However, the
following requirements apply:
Blood should be drawn from a fasting donor with a normal
hematocrit and a normal MCHC (calculated from hemoglobin and
hematocrit values determined using reference methods) and be free
of specific interferences which degrade the accuracy and/or precision
of the alternative comparative method or the i-STAT method.
• Calculation of results must correct for any systematic bias between
the reference microhematocrit method and the alternative
comparative method selected.
Reference Method
CLSI recommends that the blood samples anticoagulated with Na
2
EDTA
or K
2
EDTA be used for the microhematocrit method.* However, EDTA will
interfere with the electrolyte measurements which are used in the calculation
of hematocrit results on the i-STAT System.
*
CLSI. Procedure for Determining Packed Cell Volume by the Microhematocrit
Method; Approved Standard– Third Edition. NCCLS document H7-A3 (ISBN
1-56238-413-9). CLSI, 940 West Valley Road, Suite 1400, Wayne, Pennsylvania
19087-1898 USA, 2000.
VERIFICATION PROCEDURE FOR ACT
See Technical Bulletin: i-STAT Celite and i-STAT Kaolin ACT Heparin Linearity
Procedure

Table of Contents

Other manuals for Abbott i-STAT 1

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Abbott i-STAT 1 and is the answer not in the manual?

Abbott i-STAT 1 Specifications

General IconGeneral
BrandAbbott
Modeli-STAT 1
CategoryMeasuring Instruments
LanguageEnglish

Related product manuals