15-10 Art: 714377-00K Rev. Date: 31-Jan-12
Troubleshooting Cartridge Tests
See Troubleshooting Out-of-Range Results for Cartridges paragraph in the Performing Control Tests
on Cartridges section of this manual.
PROCEDURE FOR CARTRIDGES
Initiating Calibration Verification tests from the Quality Tests menu allows results to be stored in
seperate categories for the purpose of documentation and review. Also, measurement limits are not
applied to results in this test path. Therefore, results below and above the measurement ranges will
be reported.
A level surface includes running the handheld in the downloader/recharger.
7. Review results