5
Art: 730078-00A Rev. Date: 26-Oct-11
5. If the location where this handheld is assigned has a check mark under Use Default Profile,
under the Default customization profile: column, double click the alphanumeric code under
Preferences. Otherwise, double click the alphanumeric code under the Preferences column for
the specific location to which this handheld is assigned.
6. Once the Preferences window opens, click on the Cartridge QC tab.
7. Click on the Liquid QC Settings at the bottom of the screen.
8. In the Control Pass/Fail Determination section, click the Automatic via EVAS radio button.
Choosing this option indicates that the handheld will automatically determine whether the liquid
QC run passed or failed, based upon QC ranges contained on an electronic Value Assignment
Sheet (eVAS) file downloaded into the i-STAT 1 handheld.
Note 8.1: The Manual Control Pass/Fail Determination feature is also customizable through the
i-STAT 1 handheld keypad.
1. Power on the i-STAT 1 handheld and press MENU to get to the Administration Menu
2. Press 4 – Customization
3. Press 2 – Change
4. Type in your password and press ENT (If no password is set, just press ENT)
5. Press 4 – QC Tests
6. Press 2 – Cartridge QC
7. Press 1 to set Pass / Fail Method

Table of Contents

Other manuals for Abbott i-STAT 1

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Abbott i-STAT 1 and is the answer not in the manual?

Abbott i-STAT 1 Specifications

General IconGeneral
BrandAbbott
Modeli-STAT 1
CategoryMeasuring Instruments
LanguageEnglish

Related product manuals