Art: 723270-00A Rev. Date: 20-Apr-12
6
There are four 1.7 mL glass ampules of each level in the set.
Reactive Ingredients for TriControls Materials
Analyte
Calibration
Verification
Level 1
Calibration
Verification
Level 2 and
Control
Level 1
Calibration
Verification
Level 3 and
Control
Level 2
Calibration
Verification
Level 4 and
Control
Level 3
Calibration
Verification
Level 5
Na (mmol/L)
97 118 124 150 159
K (mmol/L)
2.30 3.00 4.00 6.30 8.20
Cl (mmol/L)
67 76 94 119 134
Glu (mg/dL)
595 285 160 65 53
Urea (mg/dL)
114 44 8.4 4.6 3.0
iCa (mmol/L)
0.40 0.90 1.35 1.58 2.40
Lac (mmol/L)
17.7 8.30 3.00 1.63 1.52
Crea (mg/dL)
15.6 4.65 1.59 0.65 0.55
PCO
2
(mmHg)
96 65 40 26 12
PO
2
(mmHg)
40 63 120 163 500
H
+
(pH)
6.550 7.025 7.390 7.610 7.850
Storage
Refrigerated storage at 2-8 ºC (35-46 ºF) should be maintained until the printed
expiration date on the box and ampule labels.
TriControls solutions may also be maintained at room temperature (18-30 ºC; 64-86 ºF)
for up to 5 days.
Do not use TriControls solutions past the labeled expiration date on the box and ampule
labels.
Ampule Use
When using cartridges that contain sensors for pH, PCO
2
, PO
2
and ionized calcium, a
separate ampule must be used for each cartridge being tested.
Do not use residual TriControls solution that may be in a syringe, ampule or capillary
tube for additional testing of cartridges that contain sensors for ionized calcium, pH,
PCO
2
, or PO
2
. However, cartridges without these sensors may be tested with remaining
fluids if that testing is performed within 10 minutes of opening the ampule.
Best Results
For best results, ampules, cartridges and handhelds should be at the same temperature.
Before Use
i-STAT TriControls solutions require different temperature stabilization times depending
on whether or not PO
2
is to be measured. If PO
2
is to be measured, equilibrate the