22-16 Art: 714384-00D Rev. Date: 07/18/06
Procedure
1. The CDS program must be running and the external interface must be
enabled.
2. Double click the i-STAT interface icon in the system tray (next to the
clock in the lower right hand corner of the screen) to open the interface
program’s main screen.
3. Click File
Options...
4. Click the Send Priority
tab. (The other tabs will be available for viewing
only.)
5. Click the Location/Method line to prioritize.
6. Right click under the Send Priority
column and select the priority:
Normal, High, Very High, from the drop down list.
7. Click
Accept to finish.