EasyManua.ls Logo

Abbott i-STAT 1 - Page 449

Abbott i-STAT 1
658 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
Rev. Date: 01-Aug-11 Art: 714180-00K PO2 - 3
Method comparison data were collected using CLSI guideline EP9-A
4
. Arterial blood samples were collected
from hospital patients in 3cc blood gas syringes and were analyzed in duplicate on the i-STAT System and
the comparative method within 5 minutes of each other.
Deming regression analysis
5
was performed on the first replicate of each sample. In the method comparison
table, n is the number of specimens in the data set, Sxx and Syy refer to estimates of imprecision based
on the duplicates of the comparative and the i-STAT methods respectively, Sy.x is the standard error of
the estimate, and r is the correlation coefficient.* Method comparisons will vary from site to site due to
differences in sample handling, comparative method calibration and other site specific variables.
* The usual warning relating to the use of regression analysis is summarized here as a reminder: For any analyte, “if the data is collected over
a narrow range, the estimate of the regression parameters are relatively imprecise and may be biased. Therefore, predictions made from these
estimates may be invalid”.
6
The correlation coefficient, r, can be used as a guide to assess the adequacy of the comparative method range in
overcoming this problem. As a guide, the range of data can be considered adequate if r>0.975.
Precision Data (mmHg)
Aqueous Control Mean SD %CV
Level 1 65.1 3.12 4.79
Level 3 146.5 6.00 4.10
Method Comparison (mmHg)
Radiometer
ABL500
Radiometer
ABL700
Bayer 845
n 45 29 30
Sxx 3.70 2.04 3.03
Syy 2.78 2.64 3.28
Slope 1.023 0.962 1.033
Int’t -2.6 1.2 -2.9
Sy.x 2.52 3.53 3.44
Xmin ---- 39 31
Xmax ---- 163 185
r 0.996 0.990 0.996
Factors Affecting Results*
Exposure of the sample to air will cause an increase in PO
2
when values are below 150 mmHg and a
decrease in PO
2
when values are above 150 mmHg (approximate PO
2
of room air).
Standing anaerobically at room temperature will decrease pH at a rate of 0.03 per hour, will increase PCO
2
by approximately 4 mmHg per hour and will decrease PO
2
at a rate of 2–6 mmHg per hour.
6
Do not ice samples before testing – PO
2
results may be falsely elevated in cold samples. Do not use a cold
cartridge – PO
2
results may be falsely decreased if the cartridge is cold.
*It is possible that other interfering substances may be encountered. These results are representative and your results may differ somewhat due to
test-to-test variation. The degree of interference at concentrations other than those listed might not be predictable.
Factors Affecting Calculated Results
sO
2
values calculated from a measured PO
2
and an assumed oxyhemoglobin dissociation curve may differ
significantly from the direct measurement.
3

Table of Contents

Other manuals for Abbott i-STAT 1

Related product manuals