System Messages
74 | English (USA) 2019-08 901418-003,01 ACIST | CVi® System User's Guide
8 Troubleshooting
Message Recommendation
Check Syringe Valve
Remove syringe assembly from
chamber…OR Clean syringe valve
sensor and check cable connection
e system has detected a syringe
in the chamber during startup
calibration. If there is a syringe in the
chamber, remove it to continue the
setup procedure. Failure to remove
the syringe will result in damage that
makes the syringe unusable.
If there is not a syringe in the
chamber, the syringe valve sensor
may need cleaning or may be
blocked.
Check Syringe Valve
To reuse syringe, turn power OFF
and ON again and press RESUME…
OR Press OK to allow removal of
syringe from chamber
e RESTART option was selected
while a syringe was already in the
chamber. To start a new case, remove
all components.
Communication Error Has Occurred Try cycling power to the system.
Communication Timeout! Try cycling power to the system.
Contrast Empty e system was unable to ll the
syringe from the contrast container.
If the contrast container is empty,
replace it with a new container. If
there is contrast remaining in the
container, make sure that the white
slide clamp is open, that there are
no air bubbles in the spike, that the
contrast tubing is secured in the
contrast sensor, and that the contrast
sensor is clean.
Counter/Position Mismatch ere may be a problem with the
components that drive the syringe
ram. If the problem persists, contact
a service representative.
Disarm e connected X-ray imaging
system has disarmed the CVi system.
Consult the X-ray imaging system's
instructions to determine how to
correct the problem.
DISARMED! Arming and disarming the system
is controlled by the X-ray imaging
system. Consult the X-ray imaging
system's instructions to determine
how to correct the problem.
DOS unable to stop injection motor
error!
e system has experienced a
hardware problem. Contact a service
representative.