ACIST | CVi® System User's Guide 901418-003,01 2019-08 English (USA) | 75
System Messages8 Troubleshooting
Message Recommendation
Error n Try cycling power to the system.
If the problem persists, contact a
service representative and report the
error number.
Forward Limit (or Forward Stop)
Reached
e syringe ram has reached its
maximum forward position. Try
pressing START REFILL to initiate a
rell procedure.
Function Timeout! e system experienced an error
during calibration. Try cycling power
to the system. If the problem persists,
contact a service representative.
Hand Controller Disabled (or HC
Disabled)
e connected X-ray imaging system
has disabled the hand controller.
Injections are controlled remotely
from the X-ray imaging system.
(Note: Philips X-ray imaging systems
may still require the hand controller
to be pressed to initiate an injection.)
Hand Controller Calibration Failed e hand controller is a single use
component and must be calibrated
for every case. Check the following
items if the hand controller
calibration fails:
e hand controller luer
connections are tightly connected
to the control panel.
e (C) button on the hand
controller was fully depressed.
e (C) button was not pressed
within 4 seconds of starting the
hand controller calibration.
e (C) button was pressed before
pressing OK on the control panel.
Hand Controller Requires
Calibration
e INJECT button was pressed
before the hand controller was
calibrated. Calibrate the hand
controller before attempting to arm
the injector.
Injection Motor Malfunction
(OS/OT)
Make sure that the stopcock is open
and that ow is not restricted in any
part of the system or patient kit. If
there are no ow restrictions and
the error persists, contact a service
representative.
Interprocessor Ping Error Try cycling power to the system.