System Messages
76 | English (USA) 2019-08 901418-003,01 ACIST | CVi® System User's Guide
8 Troubleshooting
Message Recommendation
Invalid Calibration Data e system detected a problem
with the startup calibration. Press
OK to retry the calibration or try
cycling power to the system. If the
problem persists, contact a service
representative.
Invalid Calibration Data. Can't
Resume!
When trying to resume a previously
interrupted case, the system detected
a problem with the calibration
data. Press OK to continue, or try
cycling power to the system. If the
problem persists, contact a service
representative.
Invalid HC16 Response! Try cycling power to the system.
Manifold Valve Closed e system displays this alert when
it detects a manifold in the manifold
sensor during startup calibration.
If there is a manifold in the sensor,
remove it to continue the setup
procedure.
Manifold valve failed to close n times
using current syringe.
e manifold is designed to prevent
reverse ow by automatically closing
when uid ow stops. Replace the
syringe if:
• Patient bodily uids are observed
in the patient tubing.
• Both the Manifold Valve
Open and Syringe Valve Open
indicators are on and any error
messages are displayed.
• If the system displays both the
Manifold Valve Open error and
Syringe Valve Open indicator.
Manifold Valve Open e manifold valve opens during
an injection. erefore, this alert
sometimes appears for informational
purposes. If the alert persists aer an
injection is complete, check for the
following issues:
• Make sure that the manifold is
properly seated in the sensor.
• Try purging.
• Press the Standby button to
relieve back pressure on the
manifold.
• Make sure that the white slide
clamp on the contrast tubing is
open.
• Replace the patient kit.
Motor Speed Incorrect Try cycling power to the system.